Deformability results 1 parameters explanation
The following result parameters are given in Results file and printed report:
Parameter
|
Description
|
Start time
|
Actual time that was valid after setting a new shear stress and immediately following the image stabilisation time.
|
End time
|
Actual time that was valid after the last EI measurement at the given shear stress was performed.
|
Shear stress (Pa)
|
Applied shear stress.
|
Coordinate count or Crd cnt
|
Number of coordinates in the iso-intensity curve that describes the shape of the diffraction pattern.
|
A (pixels)
|
Vertical axis of ellipse.
|
B (pixels)
|
Horizontal axis of ellipse.
|
EI
|
Elongation index.
|
Eof (pixels)
|
Error of fit. This parameter represents the average distance (in pixel units) of the iso-intensity pixels to the best-fit ellipse that resulted. It reflects how well the iso-intensity coordinates comply with the ellipse model.
|
#Tries
|
Number of diffraction patterns that were analyzed at the current shear stress.
|
#Valid
|
The number of diffraction patterns that were within the ±2SD confidence interval. This number agrees with the desired count as given in the Settings menu (<Settings<Determinations per shear stress).
|
|