First Start-up

Steady state

After switching on, the LoRRca MaxSis needs 5 minutes to warm up.

Before the instrument can be operated, you have to carry out the following checks and procedures:

  • Check that there are no foreign objects (tools, for example) in the instrument.

  • Switch ON the instrument.
  • Start-up the PC.
  • Choose LoRRca software program.
  • Select <Hardware check>.
  • Move the Bob down.

main_screen320x238

  • Choose <Camera>
  • Set camera gain at 250.
  • Close the hood.

HDW

  • Check the image:
    • OK: The instrument is now ready.

camera_alligned

  • Not OK: Misaligned
  • Check position of the projection screen.

camera_misalligned

camera_no_picture

  • For basic version choose <LoRRca IO><Rinse cup>.
    The tubes are being filled.
  • Dry BOB and CUP. Do not scratch the cup!
  • For Osmoscan version choose <LoRRca IO><Prime all>.
    The tubes are being filled.
  • Dry BOB and CUP. Do not scratch the cup!

hwc

Alignment of the projection screen:

  • Use the hexagon key 2mm to loosen the projection screen holder.

 

alignment_projection_screen

 

  • Align the projection screen until the black dot on the screen is aligned with the laser spot.
  • Tighten the screws of the projection screen holder.

camera_alligned

After this procedure, the instrument is in the start position.

See Also

Installation

Transport

Ambient conditions and setting up the instrument

Checking LoRRca functionality