Trouble shooting table

Field

Description

Check/Remedy

Waste container

Container indicates waste full when it is not. Waste liquid is clinging to the inside of the bottle and giving a false reading.

  • Flush bottle thoroughly with hot water or replace the bottle.

Power

No power

  • Check the fuses at the main power inlet.
  • Check cables
  • Replace if necessary.
  • Check the fuse on the LoRRca PC main board

Camera

No camera image

  • Check Fire wire cable. Reconnect if necessary.
  • Check Camera driver is not corrupt.

Camera

Disturbed image

Fluid flow

There is no fluid flow or not enough fluid flow in one ore more tubes

  • Check the liquid levels in the bottles
  • Check the functioning of the pinch valve, in case of stitching: perform a Prime all

In case of not correct screen, not sharp and black dot is not in the middle, the laser have to be replaced or adjusted.

To check the connection on the camera Fire wire, the handle on the right side has to come off to check if the connection is good.

See Also

Trouble shooting

Error messages

Service request protocol