Rinse and waste pump tube replacement

Important:
The rinse- and waste peristaltic pumps use different tubing, the waste pump uses QWLV090011 (3,2mm tube) and the rinse pump uses QWLV090012 (1,6mm tube) Always replace the correct tube!

The pumps are in the foot of the LoRRca MaxSis, push on the lid to open the pump drawer.

The pump tube needs to be replaced every 6 month.

  • Switch the instrument OFF.

Pump tube

pump_tube_replacement

  • Remove the old tube from the pump as shown:
  • Pull the tube from the connectors and install the new tube in the reverse order.
  • Check that all connecting tubes and the tube cover are attached.

See Also

Waste and rinse pump tube replacements

Pump connection tubing