WI-257 Daily Maintenance
Work instruction Number 257
|
Page 1 of 1
|
Purpose: Daily maintenance
|
Safety: Bio Hazard area
|
|
Instrument: Interrliner
|
Revision: 001, 2019
|
Prepare disinfectant: (if not already prepared). This disinfectant is for cleaning of all external parts that are exposed to blood.
- Go to tab and perform the procedure.
- Check system for leakage.
- Inspect the peristaltic pump tubes and connections for leaks.
- Check that liquid does not run back after the pumps have stopped.
- Clean the outer needle with disinfectant
- Check tubing from the syringe for trapped air bubbles.
- Check Diluent syringe for trapped air bubbles.
- If trapped air bubbles are found, go to tab click button and perform the function.
- Wipe outer surface and stainless steel plate below the pipettes with disinfectant.
- Check (or replace if needed) two drip trays (ESRI010318)
|