WI-208 Sample probe Interrliner

Work instruction Number 208

Page 1 of 1

Purpose: Sample probe or outer needle replacement

Safety: Bio Hazard area

 

Instrument: Interrliner

Revision: 002, January 2019

PDF-version

Lift the robotic protection hood.

Needle exchange:

BiohazardSharp warning

  1. Unscrew sample probe manually.
  2. Mark each tube for easier reconnecting to the correct nipple.
  3. Disconnect the tubes from the outer needle.
  4. Pull the sample probe, complete with outer needle, towards the front edge of the ESR analyzer unit.
    The outer needle must be supported to protect it from falling.
  5. Slide the new sample probe into the (new) outer needle.
  6. Make sure the Sample probe has a (new) O-ring QWLV050003.
  7. Install (new) sample probe ESRI050909 together with the (new) outer needle ESRI050950.
  8. Tighten the sample probe.
  9. Replace the correct tubes on the outer needle.
  10. Close the robotic protection hood.