WI-251 Fill and Clean Interrliner
Work instruction Number 251
|
Page 1 of 2
|
Purpose: Fill and Clean
|
Safety: Bio Hazard area
|
|
Instrument: Interrliner
|
Revision: 001, 2019
|
PDF-version
Applicable for Software version 6.12 and up.
Starrsed Cleaning Agent preparation Interrliner ESR analyzer unit: Fill and clean: This cycle takes about 90 minutes.
- Fill the clean adapter EHST110907 with hot de-ionized water. (+/- 150 ml, 80 °C)
- Add 15 ml Starrsed Cleaning Agent (QRR 010905) to the hot water in the adapter.
- Place the cap on the adapter and mix well.
- Put the adapter with cleaning solution on the lower tube holder.
- Select tab, , button .
|
|
Start Fill and clean procedure:
- Select button OK.
- The needle goes down and the process is started.
- When all the pipettes are filled, the needle goes back to the home position.
- After the Fill & clean process is finished, a notification is given to stop the procedure.
- Remove the adapter from the needle assembly and press "Stop".
The instrument is ready for normal operation.
|