Small air bubbles rising in pipette

AirBubbles02

Usually this is caused by a dirty or damaged fill nozzle.

  • Observe the maintenance schedules.
  • Clean the fill nozzle.
  • Check the fill nozzle for damage. If necessary, replace the fill nozzle.

Sample tube is leaking on the fill nozzle side.

  • Replace the silicon sample tube

Small air bubbles result in ESR Error 3.

See Also

Air bubbles

Foam in column

Pipette looks like zebra crossing

One air bubble about 5 mm under meniscus

One air bubble rising in pipette

Random air bubbles in pipette