One air bubble rising in pipette

AirBubbles02

  1. Usually this is caused by a wet or dirty fill nozzle.
    • The blood column should not reach right to the base of the pipette. There must be a clear air gap of 4...5mm at the bottom of each pipette.
  2. Insufficient sample volume.
    • Need more blood in the sample tube.

One air bubble rising can result in ESR Error code 3.

See Also

Air bubbles

Foam in column

Pipette looks like zebra crossing

One air bubble about 5 mm under meniscus

Small air bubbles rising in pipette

Random air bubbles in pipette