Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Report 60-Minute mode

+ REPORT EXAMPLE +(Not to scale)

-- StaRRsed--

Date 25/04/2019

Time:

15:28 

 

 

 

 

 

 

 

 

 

 

 

1

2

3

4

5

6

7

8

9

10

905001

 

84

75

CLEAR

17

60

23

 

EDTA

905002

 

14

13

Hazy<10mm

18

60

23

 

EDTA

905003

 

22

21

Hazy<25mm

19

60

23

 

EDTA

905004

 

67

61

Hazy>25mm

20

60

23

 

EDTA

905005

 

 

 

CLEAR

21

60 

23 

 

EDTA

905006

 

5

5

CLEAR 

22

60

23

 

EDTA 079

905007

 

 

 

 

24

60

23

Too many borders found

905008

 

 

 

 

25

60

23

L_err(---/ 84/ 75/200)

EDTA

905002/905003/905004
Sample results with hazy aspect

905005:
Sample result with a manual aspect, where the manual aspect is shown as a number 3 in column 6 of this data record sample.

905006:
In this sample, the dilution rate has a dilution failure of 21% and that is printed as EDTA 079.

905007
Sample results with a text error. This sample gives Too many borders found. Result of a pipette possibly filled with air bubbles.

905008
Sample result with a text error. This sample is given limit error L_err(---/ 84/ 75/200)

See Also

Result Printout

Report 30 Minute mode