QC Error messages

The general ESR errors and warnings are also applied on the QC results, see "ESR Error and Warning code messages"

When the result is within range, no message is shown.

When the result is out of range an error message is shown in the status line of the Sample screen and the QC icon is blinking on the Sample screen. When the sample mode is started again by the operator, the following messages appears:
Last QC result was out of range! Continuing could produce incorrect results! Do you still want to continue?
Press "Accept" to continue sampling without performing a new QC, press "Cancel" to return and take appropriate action.

Messages when the general setting "Temperature Correction" is switched ON:

  • "E116: QC is out of acceptable range!"
    The Sample mode is switched OFF automatically. Remaining filled pipettes are processed in the normal manner.

Messages when the general setting "Temperature Correction" is switched OFF:
The software always calculates a temperature corrected result because only temperature corrected results can be compared with the Assay mean value.

  • "E116: QC is out of acceptable range!"
    The uncorrected and the corrected result are out of range.
  • "E117: Uncorrected QC result is out of acceptable range, but corrected result is within range!"
    The uncorrected result is out of range, but the corrected result is within range.
  • "E118: Uncorrected QC result is within acceptable range, but corrected result is out of range!"
    The uncorrected result is within range, but the corrected result is out of range.

See Quality control troubleshooting and QC Results screen for more details.