Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

ESR Error and Warning code messages

ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10.
Error: No ESR result is given
Warning: Result should be reviewed before release.

The following codes are defined:

0

No errors

 

 

1

No cells/plasma found

Error

No contents could be detected in the pipette.

2

ESR Probably > 140 mm

Error

Extremely high ESR value.

3

Too many borders found

Error

More than three borders found, possibly air bubbles. See Section Troubleshooting Air bubbles.

4

Column height <nnn>

Warning

Column height must be between 180 and 210mm. <nnn> = the actual column height.

5

Measure error

Warning

The down count is not equal to the up count from the measure head.

6

Bubbles on top

Warning

Air bubbles on top of the ESR. See Section Troubleshooting Air bubbles.

7

Limit error

Error

One of the following limits are out of the setting range:

  • ESR Time
  • Column height
  • Dilution
  • Bubbles on top
  • Hazy aspect
  • Temperature

 

See Also

ESR Error