Instrument models
|
Model name Interrliner XN1 Interrliner XN2 Interrliner XN3
|
Catalogue number (REF) EHST109421 EHST109422 EHST109423
|
ESR method
|
Westergren method ICSH J. M. Jou; Int. Journal of Laboratory Hematology 2011; 33: 125-132 CLSI Procedures for the Erythrocyte Sedimentation Rate Test; Approved Standard- Fifth Edition H02-A5, Vol. 31 No. 11; 2011
|
Temperature compensation method
|
R.W. Manley: J. clin Path (1957), 10, 354
|
30 minute method
|
R. Rogers: Medical Laboratory World 1994
|
Allowed blood specimen types
|
For EDTA mode: Whole blood with < 1% EDTA anticoagulant.
In Citrate mode: Whole blood with 3.2% sodium citrate anticoagulant-diluent (not available in Software version 7.xx)
|
Automatic dilution
|
4 vols. blood + 1 vol. Starrsed Diluent
|
Reported result
|
mm after 1 hour
|
Throughput (100% filled)
|
single
|
double
|
triple
|
30 minute method
|
140/hour
|
240/hour
|
360/hour
|
60 minute method
|
75/hour
|
140/hour
|
215/hour
|
Reagents:
|
|
Reagents used
|
QRR 010931 Starrsed Diluent QRR 010947 Starrsed Disinfectant QRR 010933 Starrsed Saline QRR 010934 Starrsed Rinse Solution De-ionized water
|
Blood volume:
|
|
Aspirated blood volume per sample
|
1.4 ml in EDTA mode
|
Rack and tube types:
|
|
Rack type
|
Sysmex sample rack (low profile)
|
Sample tube types
|
Most commonly used brands/types. Closed tubes with concentric cap only.
|
Barcode reader:
|
|
Barcode reader type
|
CCD.
|
Reading capabilities
|
Most common barcode labels Code39, ITF, Industrial 2 or 5, CodaBar,EAN/UPC, CODE128
|
Environment:
|
|
Sound level
|
Less than 65 dBA
|
Environment temperature
|
18 - 28 °C
|
Relative humidity
|
10-90%
|
Total power specifications
|
|
Mains voltage
|
230VAC, 50/60 Hz
|
Fuse (20 x 5 mm)
|
Per ESR Unit
|
Slow blow 230V 2.5 Amp
|
|
Transport Unit
|
Slow blow 230V 1.6 Amp
|
|
|
XN1
|
XN2
|
XN3
|
Power consumption
|
Standby
|
90 VA
|
2x90VA
|
3x90VA
|
|
Maximum
|
650 VA
|
2x1000 VA
|
3x1500 VA
|
Heat output
|
Standby
|
86 Watt
|
2x86 Watt
|
3x86 Watt
|
|
Operation
|
460 Watt
|
2x460 Watt
|
3x460 Watt
|
Dimensions
|
XN1
|
XN2
|
XN3
|
Width
|
1795 mm
|
2890 mm
|
3985 mm
|
Height
|
1530 mm
|
1530 mm
|
1530 mm
|
Depth
|
1150 mm
|
1150 mm
|
1150 mm
|
Total weight (empty)
|
330 kg
|
542 kg
|
754 kg
|
Weight (components)
|
|
ESR analyzer unit
|
45 kg
|
Start pool
|
63 kg
|
End pool
|
41 kg
|
Conveyer/robot unit
|
140 kg
|
Protection hood
|
12 kg
|
Printer stand
|
11 kg
|
Back panels
|
18 kg (single), 22 kg (double, 26 kg (triple)
|
LIS connectivity:
|
|
Communication
|
Ethernet and serial, bi-directional according various protocols
|
Data storage:
|
|
Storage medium
|
Hard disk on external PC
|
Storage capacity indication
|
approx. 5 Mb per 1000 samples (results and raw data)
|
Compliance
|
|
|
Classification EU
|
IVD Class A
|
Classification USA
|
Class 1, 510 (k) Exempt
|
Compliant to
|
EU harmonized standards relevant to IVD Medical devices
|
Used materials
|