Technical specifications

Instrument models

Model name
Interrliner XN1
Interrliner XN2
Interrliner XN3

Catalogue number (REF)
EHST109421
EHST109422
EHST109423

ESR method

Westergren method
ICSH J. M. Jou; Int. Journal of Laboratory Hematology 2011; 33: 125-132
CLSI Procedures for the Erythrocyte Sedimentation Rate Test; Approved Standard- Fifth Edition H02-A5, Vol. 31 No. 11; 2011

Temperature compensation method

R.W. Manley: J. clin Path (1957), 10, 354

30 minute method

R. Rogers: Medical Laboratory World 1994

Allowed blood specimen types

For EDTA mode: Whole blood with < 1% EDTA anticoagulant.

In Citrate mode: Whole blood with 3.2% sodium citrate anticoagulant-diluent (not available in Software version 7.xx)

Automatic dilution

4 vols. blood + 1 vol. Starrsed Diluent

Reported result

mm after 1 hour

Throughput (100% filled)

single

double

triple

30 minute method

140/hour

240/hour

360/hour

60 minute method

75/hour

140/hour

215/hour

Reagents:

 

Reagents used

QRR 010931 Starrsed Diluent
QRR 010947 Starrsed Disinfectant
QRR 010933 Starrsed Saline
QRR 010934 Starrsed Rinse Solution
De-ionized water

Blood volume:

 

Aspirated blood volume per sample

1.4 ml in EDTA mode

Rack and tube types:

 

Rack type

Sysmex sample rack (low profile)

Sample tube types

Most commonly used brands/types.
Closed tubes with concentric cap only.

Barcode reader:

 

Barcode reader type

CCD.

Reading capabilities

Most common barcode labels
Code39, ITF, Industrial 2 or 5, CodaBar,EAN/UPC, CODE128

Environment:

 

Sound level

Less than 65 dBA

Environment temperature

18 - 28 °C

Relative humidity

10-90%

Total power specifications

 

Mains voltage

230VAC, 50/60 Hz

Fuse (20 x 5 mm)

Per ESR Unit

Slow blow 230V 2.5 Amp

 

Transport Unit

Slow blow 230V 1.6 Amp

 

 

XN1

XN2

XN3

Power consumption

Standby

90 VA

2x90VA

3x90VA

 

Maximum

650 VA

2x1000 VA

3x1500 VA

Heat output

Standby

86 Watt

2x86 Watt

3x86 Watt

 

Operation

460 Watt

2x460 Watt

3x460 Watt

Dimensions

XN1

XN2

XN3

Width

1795 mm

2890 mm

3985 mm

Height

1530 mm

1530 mm

1530 mm

Depth

1150 mm

1150 mm

1150 mm

Total weight (empty)

330 kg

542 kg

754 kg

Weight (components)

 

ESR analyzer unit

45 kg

Start pool

63 kg

End pool

41 kg

Conveyer/robot unit

140 kg

Protection hood

12 kg

Printer stand

11 kg

Back panels

18 kg (single), 22 kg (double, 26 kg (triple)

LIS connectivity:

 

Communication

Ethernet and serial, bi-directional according various protocols

Data storage:

 

Storage medium

Hard disk on external PC

Storage capacity indication

approx. 5 Mb per 1000 samples (results and raw data)

Compliance

 

 

Classification EU

IVD Class A

Classification USA

Class 1, 510 (k) Exempt

Compliant to

EU harmonized standards relevant to IVD Medical devices

Used materials

ROHS 2011/65/EU and 2015/863/EU