Fill procedureThe protective hood must be closed to prevent injuries from the moving parts of the robotic unit. Sysmex XN versions: Standalone versions: The Rack will be transported to the barcode manipulator to read the barcode labels. Then the rack is moved to the robotic arm and three sample tubes are picked up for mixing. After mixing, the tubes are replaced into the rack and the Rack is moved one position. The robotic arm is picking up the sample tube and places the tube into the needle unit to be aspirated. After all sample tubes in a rack are processed, the rack will be removed and transported to the End Pool. ATTENTION: It is not allowed to change sample tubes in a rack or remove sample tubes while the rack is processed. This may cause malfunction of the instrument.
|
|||||||