Cleaning waste separator

The separator is designed to separate liquid from the air and can handle a lot of blood, rinse and other used reagents from the instrument. After a period of time the separator is getting dirty and therefore it needs to be cleaned weekly.

Symptoms of a dirty separator:

  1. Separator errors.
  2. Foam in the separator.
  3. Waste pump cannot sufficiently remove the waste from the separator.

Cleaning waste separator.

Removing

17402 Remove waste separator

Cleaning

s

A. Clean all parts with hot water and a brush.
B. Use some acid free vaseline on the screw-thread of the glass jar, the top rim of the glass jar, the O-ring for the HEPA filter and the black PVC pipe on the back.

Replacing

17402 Replace waste separator

C. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002)

See Also

Weekly

Check the sensors in service mode