Display patient results

To Maintenance screen History screen To Service screen Close program To Reagents status screen To Setting screen To Sample screen Extra search options can be select Pipette data in carrousel Sample history Rack history Send patient result to Host Back to sample history list Shown raw data in the pipette Raw data shown in a graphice The patient sample identification Time and date of the sample Result corrected for the temperature. Dilution rate of the sample Room temperature at the  time of the reading of the sample. Pipette number of the processed sample Error code when error is found during the measurement like Too many borders found. Input patient  indent and press Search  to find patient ID in the database. Analyse results ESR Statistics QC results

In the window Select date to show results: double click on the file name to select the results of the selected date.
Press Refresh to refresh the list of available files.
In the window Sample ID type the sample ID information and press Search.

Press Options for the following search options:

  • Show today's results.
  • Show today's results from a selected time frame of the day.
  • Show results of a number of past days. Default value is set for 7 days.
  • Show results of a specific day.
  • Show results of the range between the first selected date to the next selected date.

From the selected Sample ID detailed information is shown on this screen.

Sample ID

Sample Identification number

Aspect code

Shows the aspect code (e.g. Hazy <10)

ESR 30 min

The 30 minute method is used. This is the measured 30 minutes value.

ESR 60 min

When the 60 minute method is used, this is the measured 60 minutes value.
When the 30 minutes method is used, this is the calculated 60 minutes value.

ESR 60 min T.Corr

Temperature correction is used. This is the 60 minutes value corrected to 18°C.

Date / time

Date and time of the measurement of the result.

ESR time (min.)

Actual duration of the ESR.

Dilution %

The calculated dilution rate after aspiration of the sample.

Temperature (°C)

Room temperature at the time of the measurement of the sample.

Pipet number

Pipette in which the sample was measured.

Error code

Shows any ESR error or warning code (e.g. "Too many borders found").

A complete overview of measurement data is shown with activating the option "Display raw data".

See Also

Display Sample history