Check sensors

Check sensors explanation After measuring, the results are displayed graphical and numerical Program is closed via the  shut down procedure Service screen To Sample screen To History screen Maintenance history To Setting screen To Reagents status screen Rack unit setting Serial output setting To  Maintenance screen

When there is a sensor alarm, an alarm indicator is shown in the tab Maintenance.

All functions to check the status of the sensors are grouped under button Check sensors.

  • Check Fill stop sensor: Click the Check button. The green light is shown if the sensor value is in range.
  • Check temperature sensor: Value must be equal to the actual room temperature near the pipette belt.
    The value can be set in tab Settings.
  • Check Diluter start sensor: This sensor is only used in EDTA mode. If the diluter does not start during the aspiration, the status of this sensor must be checked.
    Click the Check button. The green light is shown if the sensor value is in range.
  • Check Diluent flow sensor: This sensor is only used in EDTA mode. When activated, the LED Down is green and the LED Up is red. When the button Test is clicked, the LED Up must become green. After finishing the test, both LED's must be green.
  • Check Separator sensor: Click the Check button. The green light is shown if the sensor value is in range.
  • Check Flow sensor: Click the Check button. The green light is shown if the sensor value is in range.
  • Check Measure sensor:
    Clean sensors first before executing this function.
    Click the Check button. The green light is shown if the sensor value is in range.
    Press the button Measure. The pipette currently at the measure position will be measured.
    The results are displayed in graphical form:

take out pipette bza

Measure head start position correct

take out pipette bza

Measure head start position wrong

When a test pipette is installed at the measuring position the result of the test pipette is displayed in the field "ESR (mm)".

Loep

When the sensor is out of range and the light is red, the sensor values can be checked by turning on the service mode.

 

In This Section

Explain check sensor screen short

Check sensors in service mode

Check sensor functions overview

Fill stop sensor

Temperature sensor

Diluter start

Diluent flow sensor

Separator sensor

Flow sensor

Measure sensor