Saline tubing pops off from outgoing saline pump connector (FAQ)

Symptom :

Saline tubing pops off from outgoing saline pump connector.

Hint :

  1. Debris from the silicon pump tubing is blocking the outer needle port.
    • Clean the outer needle port and replace the pump tubing ESRI090903

Additional information :

Peristaltic pump tubes replacement (WI-162 Peristaltic pump tube replacements)

Applicable

Interrliner
Starrsed RS
Starrsed Compact

FAQ 30068

See Also

Reagents

Reagent container full not detected (FAQ)

Poor or insufficient Saline flow (FAQ)

No flow of deionized water (FAQ)