Dilution error detection 0 till 25%
  • Dilution error detection 0 till 25%
    Dilution Error deviation report. If a dilution error occurs during the aspiration sequence, an audible alarm sounds and the deviation value will be shown on the screen. When the measure unit has evaluated the sample, the deviation value will be printed after the text "EDTA"
    Sending or not sending results with dilution errors to the output is optional, see Limit error settings (Limit error settings, Limit error settings).

    Example: Dilution error detection is set at 10%. When the dilution error is outside the 10% range, in the last column of the report EDTA 079 or EDTA 121 is printed which indicating this sample is 21% under or over diluted.

See Also

Diluter settings

Dilution settings explain

Dilution adjustment 60 till 140%

Auto dilution adjust

Diluent flow check