Flow sensor
  • Check Flow sensor: The vacuum unit switches on and the values must be in this range:
    Flow: 0980 ± 60 Abs: 0320 ±10
    Note: If for example the yellow orifice is blocked the flow will be: Offset: 0050 ±5.