Needle not in top position

E4

E4 is the error code and the text in the status line Sample probe not in top position ! (home). Piercing needle returns not at the home position after sampling a tube.

  • Check the home sensor.
  • Sample probe motor is faulty.
  • Sample probe motor driver on needle board is faulty.
  • Sample probe is blocked
  • Fatal error, call distributor.

E14

E14 is the error code and the text in the status line Outer needle motor position error ! (home). Outer needle did not reach the home sensor within a certain time limit.

  • Check home (top) sensor.
  • Outer needle (tube) motor is faulty.
  • Outer needle motor driver on needle board is faulty.
  • Motor is blocked.
  • Outer needle is blocked.
  • Fatal error, call distributor.

See Also

Needle Systems