WI-232 Weekly maintenance
Work instruction Number 232
|
Page 1 of 4
|
Purpose: Weekly maintenance
|
Safety: Bio Hazard area
|
|
Instrument: Starrsed TL
|
Revision: 002, 2019
|
PDF-version
Change instrument status in the LAS from "online" to "offline"
Clean Fill nozzle
1.
|
|
2.
|
|
3.
|
|
4.
|
Clean fill nozzle
The use of a toothbrush and detergent is recommended.
- Carefully scrub the fill nozzle inner part.
- Use a tissue to dry the fill nozzle.
|
5.
|
|
6.
|
|
7.
|
|
Clean Liquid separator
Removing
A. Clean all parts with hot water and a brush. B. Use some acid free vaseline on the screw-thread of the glass jar, the top rim of the glass jar, the O-ring for the HEPA filter and the black PVC pipe on the back.
Cleaning
C. If applicable replace the bacterial HEPA filter (For Maintenance Level 4: Exchange filter QWLV040002)
Replacing
Check sensors
Vacuum pressure check
- Go to tab -> . Select box.
Flow: 0980 ± 60 Abs: 0320 ±10 If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
- Go to tab -> . Select box.
Fill stop sensor FS 90..140..165
Diluter Start sensor check
- Go to tab -> . Select box.
Diluter start sensor 400-550-700
Measure sensor check
- Go to tab -> Selectbox.
Measure sensor MS 50 ±10
Temperature sensor check
- Go to tab -> Selectbox.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
- Go to tab -> Selectbox.
Press test. When test is finished, signal Standby and Flow must shown as activated.
Separator check
- Go to tab -> Selectbox.
Separator sensor <200 600 >700
Liquid sensor check
- Go to tab -> Selectbox.
Check liquid sensor: Press test. When test is finished, the OK signs are shown at Rinse and Drain if the liquid sensor is active.
|
Check and clean instrument
Prepare disinfectant: (if not already prepared). This disinfectant is for cleaning of all external parts that are exposed to blood.
- Go to tab and perform the End-of-day wash procedure.
- Open cover and front doors.
- Check system for leakage.
- Check drip trays under pump assembly and needle assembly .
- Open carousel plate and inspect the peristaltic pump tubes and connections for leaks.
- Check tubing from the syringe for trapped air bubbles.
- Check Diluent syringe for trapped air bubbles.
- If trapped air bubbles are found, go to tab click button and perform the function.
- Clean needle top with disinfectant.
- Check if the suction tube in the guide fork is free of clogging, remove clogging if needed.
- Wipe stainless steel plates below the pipettes with disinfectant.
- Close cover/front doors and activate the instrument in the LAS.
|