Instrument model
|
Model name
|
|
Catalogue number (REF)
|
Starrsed TL
|
230V
|
TLAX109100
|
Starrsed TL
|
110V
|
A0026352
|
ESR method
|
Westergren method ICSH J. M. Jou; Int. Journal of Laboratory Hematology 2011; 33: 125-132 CLSI Procedures for the Erythrocyte Sedimentation Rate Test; Approved Standard- Fifth Edition H02-A5, Vol. 31 No. 11; 2011
|
Temperature compensation method
|
R.W. Manley: J. clin Path (1957), 10, 354
|
30 minute method
|
R. Rogers: Medical Laboratory World 1994
|
Allowed blood specimen types
|
In EDTA mode: Whole blood with < 1% EDTA anticoagulant.
In Citrate mode: Whole blood with 3.2% sodium citrate anticoagulant-diluent (not available in Software version 7.xx)
|
Automatic dilution
|
4 vols. blood + 1 vol. Starrsed Diluent
|
Reported result
|
mm after 1 hour
|
Throughput:
|
|
30 minute method
|
135/hour
|
60 minute method
|
75/hour
|
Reagents:
|
|
Reagents used
|
QRR 010931 Starrsed Diluent QRR 010947 Starrsed Disinfectant QRR 010933 Starrsed Saline QRR 010934 Starrsed Rinse Solution De-ionized water
|
Blood volume:
|
|
Aspirated blood volume per sample
|
1.4-1.6 ml in EDTA mode
|
Track and tube types:
|
|
Track types
|
Inpeco FlexLab (Inpeco, Siemens Aptio, Abbott ), GLP Systems Track
|
Sample tube types
|
Brands/types as recommended by track supplier. Closed tubes with concentric cap only.
|
Barcode reader:
|
|
Barcode reader type
|
Image based
|
Reading capabilities
|
Most common barcode labels Code39, ITF, Industrial 2 or 5, CodaBar,EAN/UPC, CODE128
|
Total power specifications
|
|
Mains voltage
|
115/230V, 50/60 Hz
|
Fuse (20 x 5 mm)
|
Slow blow 230V, 4 Amp / 115V, 6.3 Amp
|
Power consumption
|
Standby 60 VA / Operation 500 VA
|
Heat output
|
Standby 80 Watt / Operation 800 Watt
|
Environment:
|
|
Sound level
|
Less than 65 dBa
|
Environment temperature
|
18 - 35 °C
|
Relative humidity
|
10-90%
|
Total dimensions
|
|
Minimum Working space
|
1700x1100 mm
|
Total height (cabinet)
|
1340 mm
|
Weight empty
|
140 kg
|
LIS connectivity:
|
|
Communication
|
Ethernet (TCP/IP), bi-directional according various protocols
|
Data storage:
|
|
Storage medium
|
Hard disk on external PC
|
Storage capacity indication
|
approx. 5 Mb per 1000 samples (results and raw data)
|
Compliance
|
|
|
Classification EU
|
IVD Class A
|
Classification USA
|
Class 1, 510 (k) Exempt
|
Compliant to
|
EU harmonized standards relevant to IVD Medical devices
|
Used materials
|
ROHS 2011/65/EU and 2015/863/EU
|