ESR Error and Warning code messages
ESR "ERROR" and "WARNING" code messages. This code appears in the "sample data record" at column 10. Error: No ESR result is given Warning: Result should be reviewed before release.
The following codes are defined:
0
|
No errors
|
|
|
1
|
No cells/plasma found
|
Error
|
No contents could be detected in the pipette.
|
2
|
ESR Probably > 140 mm
|
Error
|
Extremely high ESR value.
|
3
|
Too many borders found
|
Error
|
More than three borders found, possibly air bubbles. See Section Troubleshooting Air bubbles.
|
4
|
Column height <nnn>
|
Warning
|
Column height must be between 180 and 210mm. <nnn> = the actual column height.
|
5
|
Measure error
|
Warning
|
The down count is not equal to the up count from the measure head.
|
6
|
Bubbles on top
|
Warning
|
Air bubbles on top of the ESR. See Section Troubleshooting Air bubbles.
|
7
|
Limit error
|
Error
|
One of the following limits are out of the setting range:
- ESR Time
- Column height
- Dilution
- Bubbles on top
- Hazy aspect
- Temperature
|
|