Check the sensors in service mode
Vacuum pressure check
- Go to tab -> . Select box.
Flow: 0980 ± 60 Abs: 0320 ±10 If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
- Go to tab -> . Select box.
Fill stop sensor FS 90..140..165
Diluter Start sensor check
- Go to tab -> . Select box.
Diluter start sensor 400-550-700
Measure sensor check
- Go to tab -> Selectbox.
Measure sensor MS 50 ±10
Temperature sensor check
- Go to tab -> Selectbox.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
- Go to tab -> Selectbox.
Press test. When test is finished, signal Standby and Flow must shown as activated.
Separator check
- Go to tab -> Selectbox.
Separator sensor <200 600 >700
|