Appendix - Maintenance schedule
Maintenance performed periodically or after number of samples
Maintenance Schedule Starrsed TL (Example)
|
Daily Maintenance Perform End-of-day wash, Check tubing/diluent syringe, Clean instrument
|
Weekly Maintenance Perform End-of-day wash, Check tubing/diluent syringe, Clean instrument, Clean Fill nozzle, Clean waste separator, Check sensors
|
Sample volume: 450 per day (5 working days)
|
Level 4
|
Level 3
|
Level 2
|
Level 1
|
Parts
|
Total amount/year
|
Or earlier after number of samples
|
7.500s
|
23.000
|
46.000
|
91.000
|
|
91.000
|
Replace Fill nozzle O-ring
|
0
|
0
|
0
|
0
|
SQWLV050004
|
12
|
Replace bacterial air filter (HEPA)
|
0
|
0
|
0
|
0
|
SQWLV040002
|
12
|
Replace peristaltic tube assembly (2x)
|
0
|
0
|
0
|
0
|
SESRI090902
|
24
|
|
|
|
|
|
SESRI090903**
|
|
Run Fill and clean
|
0
|
0
|
0
|
0
|
|
|
Check/Clean Liquid sensor
|
0
|
0
|
0
|
0
|
|
12
|
Replace O-ring inner needle**
|
0
|
0
|
0
|
0
|
SQWLV050002
|
12
|
Replace waste pump tube
|
|
0
|
0
|
0
|
SQWLV090011
|
4
|
Replace Pinch valve tube
|
|
0
|
0
|
0
|
SESRI010246
|
4
|
Replace Blue disc filter
|
|
0
|
0
|
0
|
SQWLV040003
|
4
|
Replace Teflon tips (diluter syringe)*
|
|
|
0
|
0
|
SQWLV030901
|
2
|
Replace Fill block washer
|
|
|
|
0
|
SESRI030906
|
1
|
Replace needle assembly**
|
|
|
|
0
|
STLAX090901
|
1
|
Replace needle sealing block o-ring (2x)
|
|
|
|
0
|
SQWLV050080
|
2
|
Replace needle
|
|
|
|
0
|
SBANG040050
|
1
|
Replace Pipette valves bodies (84 pieces)
|
|
|
|
0
|
SQTST040001
|
84
|
Replace all tubing (Tubing set)
|
|
|
|
0
|
STLAX079200
|
1
|
Grease needle holder/drive spindle
|
|
|
|
0
|
Molykote EM-30L
|
-
|
*Only applicable for Diluter syringe QWLV030009, Diluter syringe QWLV030021 has to be replaced after 2 years
** Only if modification A0017825 is not executed.
Note:
Periodical Maintenance Kit STLAX110985, containing all parts for yearly maintenance
Two years Periodical Maintenance Kit: SESRI120906
|
ESRI070923
|
Fill tube assembly to regulator
|
ESRI070924
|
Fill tube assembly from regulator
|
QWLV030021
|
Diluter Syringe
|
QWLV080002
|
Quad Ring
|
|