Fill and clean procedure
Note: Each pipette on the pipette belt will be filled with Starrsed Cleaning Agent, after one hour the first pipette is washed and dried.
Fill and clean with adapter:
- Fill beaker with hot de-ionized water (± 130 ml, ±80°C).
- Add ±13 ml cleaning agent (QRR 010905) to the hot de-ionized water.
- Mix the solution well.
- Fill adapter TLAX110906 with the cleaning solution.
- Turn adapter upside down and remove cap (from dummy tube) and start filling the adapter until the dummy sample tube is filled without air. - Place cap on dummy sample tube and turn adapter again, fill the adapter completely. Check for air in the dummy sample tube. - Select Starrsed TL on the track management system and put instrument "offline"
- Ensure that the Starrsed TL is in Service mode, the status is also indicated with orange LED light.
- Enable "Sample Mode On"
- Ensure the needle is in "Down" position.
- Select ;
- Select button . Now the tube disk can be moved freely.
- Place the adapter with dummy tube on the tube disk. Ensure tubing between adapter and dummy sample tube is below the tube disk.
- Rotate the tube disk until the dummy sample tube is centered over the needle.
- Close the hood and press
- The needle goes up and the fill and clean process is started. (This cycle takes about 90 minutes).
- When all the pipettes are filled, the needle goes back to the home position.
- Open the hood and remove the adapter unit.
- Solve any outstanding errors on the track information system.
- Close the hood
- Put the Starrsed TL back online.
|