Manual home Send Feedback Print | ||||||
Limit error settingsWhen an option (at Limit error settings) is set to YES and this limit error occurs, results will be printed/send to the LIMS. Together with the sedimentation time and dilution rate (which are still printed at the usual position), the operator/analyst can see what caused the error and may or may not use the ESR values which are preserved in the error message. When an option is set to NO and this limit error occurs, the fields for 30 min ESR, 60 min ESR and the temperature corrected ESR are filled with spaces and thus results are not printed/send to the LIMS. The error message in the error field (column 10) indicates that at least one of the limits (ESR time, dilution rate, column height, bubbles on top, hazy aspects and temperature) has been exceeded. Description of the error message L_err(hhh/www/ttt/ccc) :
Example of a limit error message:
Note: If limit error filters in the LIMS are not corresponding with the limit error settings on the instrument it is possible that not all expected results are send to the LIMS.
|
||||||
To top of page | Mechatronics home Send Feedback Print | |||||
|