Level 4 Maintenance instructions
NOTE: Chip card type S is required !
Additional instructions for level 4:
- Check the waste pump tube condition. Replace the tube if it appears worn or leaking.
- Inspect sample needle condition:
Menu - - Press [ ARROW DOWN ] twice to expose the needle tip(s). If the tip(s) are worn, replace the needle.
Nozzle O-ring and pipette seal ring replacement
1.
|
|
2.
|
|
3.
|
|
4.
|
|
5.
|
|
6.
|
|
5. Clean Fill nozzle with hot water
|
Note: Replace the flat pipette seal ring BANG030052 and / or replace the nozzle O-ring QWLV050016
|
7.
|
|
Note: Check if all 5 O-rings QWLV050003 are in place and / or if applicable replace the O-rings QWLV050003
|
8.
|
|
9.
|
|
10.
|
|
|