Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Sedimentation measurement principle

The automatic reading of the Westergren sedimentation pipettes is performed by moving an optical sensor along the pipettes. While the sensor is moving, it is taking a reading every 0.25 mm. The sensor is reading the absorption of infrared light through the Westergren pipette filled with blood. From these readings, values at a number of absorption levels are determined. All absorption figures are relative to the darkest and lightest reading (darkest = 100 % and the lightest = 0 % absorption respectively).

By definition the levels are:

87.5%

Cells/ plasma separation

75.0%

Hazy detection

50.0%

Meniscus detection

ESR graph

Graphic showing typical absorption values of a sample

Although the result of the ESR measurement is expressed in a quantifiable unit ("mm"), it is not a measurement that involves measurable quantities of any analytes or substances. What is measured is the result of a physiological process that is influenced by the properties of the blood sample and external factors. The ESR measurement must therefore be regarded as a qualitative process.

See Also

Introduction

Intended purpose of device

Indication for use of device

Disclaimer

Instrument overview

Explanation of symbols

EQAS

Starrsed line of ESR instruments

Time of test

Dilution principle

To top of page Mechatronics home Send Feedback Print