ESR Error and Warning code messages
Error: No ESR result is given Warning: Result should be reviewed before release.
No errors
|
|
|
No cells / plasma found
|
Error
|
No contents could be detected in the pipette.
|
ESR Probably >140 mm
|
Error
|
Extremely high ESR value.
|
Too many borders found
|
Error
|
More than three borders found, possibly air bubbles.
|
Fill Error
|
Error
|
Air detected below blood column.
|
Column height <nnn>
|
Warning
|
Column height must be between 180 and 210mm. <nnn> = the actual column height.
|
Measure error
|
Warning
|
The down count is not equal to the up count from the measure head.
|
Bubbles on top
|
Warning
|
Air bubbles on top of the ESR.
|
Limit error
|
Error
|
One of the following limits are out of the setting range:
|
|