ESR Error and Warning code messages

Error: No ESR result is given
Warning: Result should be reviewed before release.

No errors

 

 

No cells / plasma found

Error

No contents could be detected in the pipette.

ESR Probably >140 mm

Error

Extremely high ESR value.

Too many borders found

Error

More than three borders found, possibly air bubbles.

Fill Error

Error

Air detected below blood column.

Column height <nnn>

Warning

Column height must be between 180 and 210mm. <nnn> = the actual column height.

Measure error

Warning

The down count is not equal to the up count from the measure head.

Bubbles on top

Warning

Air bubbles on top of the ESR.

Limit error

Error

One of the following limits are out of the setting range:

  • ESR Time
  • Column height

 

See Also

Analyser ERROR