Appendix - Maintenance schedule

Maintenance performed periodically or after number of samples

Maintenance schedule Starrsed ST

Sample volume: 100 per day (5 working days)

Daily

Once a week

Level 4 (monthly)

Level 3 (quarterly)

Level 2 (halfyearly)

Level 1 (yearly)

Parts

Total amount / year

 

or earlier after number of samples:

 

 

2.000 samples

6.000 samples

12.000 samples

24.000 samples

Clean outside of the aspiration needle

X

X

X

X

X

X

 

 

End-of day-wash procedure / Wash all pipettes

X

X

X

X

X

X

 

 

Check blood sensor 110-160

 

X

X

X

X

X

 

 

Measure sensor 40-60 (air value)

 

X

X

X

X

X

 

 

Inspect sample needle condition

 

X

X

X

X

X

 

 

Clean fill nozzle

 

X

X

X

X

X

 

 

Check waste tube condition

 

 

X

X

X

X

 

 

Replace nozzle O-ring and pipette sealing ring

 

 

X

X

X

X

QWLV050016/
BANG030052

12/12

Replace needle sealing block (double needle version)

 

 

 

X

X

X

BANG049002

4

Replace O-rings needle block (single needle version)

 

 

 

X

X

x

QWLV050080

2

Replace disk filter

 

 

 

X

X

X

QWLV040001

4

Replace waste pump tube

 

 

 

 

X

X

QWLV090011

2

Replace plunger tips (if needed)

 

 

 

 

 

X

BANG020003

24

Replace needle assembly (double needle version)

 

 

 

 

 

X

BANG049001

1

Replace needle (single needle version)

 

 

 

 

 

X

BANG040050

1

Replace all tubing (with tubing set)

 

 

 

 

 

X

BANG079200

1

Replace teflon tips (repair set Diluter syringe)*

 

 

 

 

 

X

QWLV030901

1

Replace quad ring diluter

 

 

 

 

 

X

QWLV080002

1

Replace O-rings fill nozzle

 

 

 

 

 

X

QWLV050003

5

*Only applicable for Diluter syringe QWLV030009, Diluter syringe QWLV030021 has to be replaced after 2 years