Starrsed ST system messages

During normal operation the following "System messages" may occur:

  1. Waiting tube
    • If a filled pipette is at the measuring position before the elapsed time has finished and the operator is ready to fill the next pipette, the Waiting tube message will be displayed.
    • To continue the sample loading sequence the operator must wait until the pipette at the measuring position has been measured.
  2. Printer failure
    • Check paper quantity.
    • Note: data will be stored in the buffer and can be printed afterwards.
  3. Reagents level empty message
    • All reagent bottles have level detectors; the display shows an error that indicates which reagent bottle(s) is (are) empty.
    • Prepare new reagent as described in section Quick start up.
  4. Waste bottle full message or No waste bottle message
    • The waste container also has a level detector. If a waste error is shown on the display, the Starrsed ST will stop filling and cleaning pipettes until a new or empty container has been installed. Preferably replace it with a cleaned one.
    • Press [ ESC ] to clear the error.
    • Check the values of the waste sensor at Menu Set waste sensor.

bio_hazerd_sign

Warning:
The waste must be treated as potentially infectious (biohazardous) material.