Starrsed ST system messages
During normal operation the following "System messages" may occur:
- Waiting tube
- If a filled pipette is at the measuring position before the elapsed time has finished and the operator is ready to fill the next pipette, the Waiting tube message will be displayed.
- To continue the sample loading sequence the operator must wait until the pipette at the measuring position has been measured.
- Printer failure
- Check paper quantity.
- Note: data will be stored in the buffer and can be printed afterwards.
- Reagents level empty message
- All reagent bottles have level detectors; the display shows an error that indicates which reagent bottle(s) is (are) empty.
- Prepare new reagent as described in section Quick start up.
- Waste bottle full message or No waste bottle message
- The waste container also has a level detector. If a waste error is shown on the display, the Starrsed ST will stop filling and cleaning pipettes until a new or empty container has been installed. Preferably replace it with a cleaned one.
- Press [ ESC ] to clear the error.
- Check the values of the waste sensor at Menu Set waste sensor.
|
Warning: The waste must be treated as potentially infectious (biohazardous) material.
|
|