Fill errors during sampling

After filling the sample into the pipette, the quality of the aspiration is checked by pre-measuring the pipette. The following parameters are checked:

  • the column height
  • the usage of diluent
  • possible air bubbles

In case of air detection at the bottom of the pipette E20 Fill Error ! is shown on the display. This error indicates dilution problems or insufficient sample flow.

If there are other deviations, a beep will sound after the check and an ESR Error or warning is reported.

See Also

Sample routine procedure

Sample Mixing

Placing sample tubes on the needle system

Aspirating the sample

Aborting the sample sequence

Automatic rinsing of the sample system