Settings parameter

Description

Range

BOB radius (mm)

Outer radius of the static bob, in millimeters.

[5, 50] mm

CUP radius (mm)

Inner radius of the rotating cup, in millimeters.

[5, 50] mm

Default medium viscosity (m×Pas) (PVP)

Default viscosity of the test suspension. This value is the proposed viscosity in the "Insert a blood sample" dialog.

[0.1, 200] mPa×s

Fill empty speed (rps)

Cup revolution speed (revolutions per second) used at the moment of inserting or removing the test suspension.

[0.1, 10] rps

Shear stress for camera adjustment (Pa)

Shear stress applied during camera adjustment (diaphragm).

[0, 100] Pa

Bob temperature (°C)

Desired bob (and hence, suspension) temperature in degrees Centigrade.

[20, 45] °C

Auto start if temp. is stable within (°C)

If temperature is within ± 0.2 °C, the measurement starts after the value given at parameter ... for at least (see next)

[0.1, 10] °C

… for at least (s)

Time to automatic start

[5, 300] s