Explain LoRRca Status

Function

Description

At home position

The LED is lit when cup is at home position (= fixed position along cup circumference where back-scatter intensity is acquired)

Intensity (au)

Back-scatter intensity (arbitrary units), should be within the range [0, 100] au during measurements (when blood is inserted).

BOB temperature

Current BOB temperature in degrees Centigrade

Run motor at (rps)

Motor control. Set to desired cup revolution speed (rps = revolutions per second).

Note: During this test, the laser is turned on (flash) in order to enable back-scatter intensity measurements.

Return to Hardware check status.

See Also

Hardware check status

Hardware check IO screen

Hardware check Camera screen