PVP solution routine

Remove the PVP-bottles after use and store at a cool and dry place (4-6 °C.)

The instrument bottles have to be cleaned after use. Rinse with hot tap water and purged with Ultra pure water. Drain the bottles as much as possible and let it dry sufficiently.

In case of a new PVP batch do not forget to change the default viscosity value* in the PC software (LoRRCA Screen Icon Settings screen/General options). The batch information can be found on the Mechatronics website.

  1. Go to www.mechatronics.nl .
  2. Select "Support" at the top.
  3. Select "LoRRca MaxSis & LoRRca I/II" -> "Viscosity PVP".
  4. Click on the link "Viscosity PVP".
  5. Log in with user name: "lorrcauser" and password: "maxsis".
  6. On the page Type in your lot number and press Search
  7. Click on "Specifications" at the article number and Lot number which is mentioned on the label.
  8. Check or change (if needed) the default medium viscosity at 'Settings screen/General options".

* For Osmoscan measurements the mean of the viscosity of the HIGH and LOW has to be set as default medium viscosity.
For the other measurements the viscosity of the isotonic PVP (used for blood preparation) has to be set as default medium viscosity.