Technical specifications
Technical specifications for the LoRRca MaxSis:
LoRRca instrument models:
|
|
|
|
|
|
Model
|
Model name
|
Catalogue number
|
|
LoRRca MaxSis
LoRRca MaxSis Osmoscan
|
LORC109000
LORC109001
|
Mains voltage
|
100/240V
|
50-60Hz
|
Fuse (20 x 5 mm)
|
Slow blow 220V
|
1.6 Amp
|
|
Slow blow 110V
|
3.15 Amp
|
Power consumption
|
Standby
|
100 VA
|
|
Maximum
|
150 VA
|
Heat output
|
Standby
|
100 Watt
|
|
Full operation
|
150 Watt
|
Dimensions
|
Width
|
470 mm
|
|
Height with closed lid
|
400 mm
|
|
Height with open lid
|
630 mm
|
|
Depth
|
300 mm
|
Total weight
|
|
70 Kg
|
Min. table size
|
Width
|
1800 mm
|
|
Depth
|
700 mm
|
Sound level
|
|
< 55 ± 2 dB
|
Working temperature
|
|
15 - 35 °C
|
Relative humidity
|
|
10 - 90%
|
Specifications of other equipment
Camera
|
Objective mount
|
C-mount
|
|
Connection
|
IEEE94
|
|
Minimal resolution
|
780x582 black/white
|
|
Dimensions
|
40x40x40 mm
|
Laser
|
Wave length
|
655 nm
|
|
Focus distance
|
170 mm
|
|
output power
|
Max 3 mW
|
Tech specs Couette assembly
CUP
|
Inner Radius
|
16.01 ± 0.005 mm
|
|
Surface roughness
|
Ra<0.026 µm
|
|
Material
|
PMMA
|
BOB
|
Outer Radius
|
15.65 ± 0.005 mm
|
|
Surface roughness
|
Ra<0.026 µm
|
|
Material
|
BK7
|
|
Ambient temperature
|
+ 4°C
|
Drive
|
Minimal cup speed
|
0.03 rps
|
|
Maximum cup speed
|
20 rps
|
|
Break time from sudden stop from 106 rpm to 0 rpm
|
<30 ms
|
Shear rate
|
|
0 - 800 s-1
|
Warming-up time Bob
|
Maximum
|
2 minutes
|
Heater
|
Heating accuracy
|
± 0.2 °C
|
|
Range
|
25 - 45 °C
|
Tech specs Gradient assembly
Gradient
|
mixer range
|
0 - 25 mS/cm
|
|
mixer accuracy
|
±-0.5 mS/cm
|
|
PVP
|
50 - 500 mosm/kg
|
|
Duration
|
maximum 10 minutes
|
|
Dilution (with full blood)
|
< 200:1 ± 20%
|
VC elongation measurement
|
|
<2% at 3Pa
|
|