Waste disposal

The waste container has a level sensor and as soon as the level sensor generates a waste error, the waste container must be emptied. The waste must be treated as potentially infectious (biohazardous) material and disposed of according to local regulations. Preferably, discard the complete waste container and replace it with a cleaned one. Clear the error.

Biohazard

If the waste line is to be connected to a centralised waste collection system, the following requirements must be met:

  1. Waste tube must not exceed 5 meters or 18 feet in length.
  2. Drain height must not be higher than the original waste container inside the instrument.

Disclaimer: Check the specifications of the central waste system for rules about discharging the waste.