Check the sensors in service mode

Vacuum pressure check

  • Go to tab Maintenance -> Check sensor. Select Check Flow sensor box.
    Flow: 0980 ± 60 Abs: 0320 ±10
    If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.

Fill Stop sensor check

  • Go to tab Maintenance -> Check sensor. Select Check Fill stop sensor box.
    Fill stop sensor FS 90..140..165

Diluter Start sensor check

  • Go to tab Maintenance -> Check sensor. Select Diluter start sensor box.
    Diluter start sensor 400-550-700

Measure sensor check

  • Go to tab Maintenance -> Check sensor. Select Check measure sensor box.
    Measure sensor MS 50 ±10

Temperature sensor check

  • Go to tab Maintenance -> Check sensor. Select Check Temperature sensor box.
    Temperature sensor TS [Room temperature]

Diluent flow sensor check

  • Go to tab Maintenance -> Check sensor. Select Check Diluent flow sensor box.
    Press test. When test is finished, signal Standby and Flow must shown as activated.

Separator check

  • Go to tab Maintenance -> Check sensor. Select Check Separator sensor box.
    Separator sensor <200 600 >700

See Also

Weekly

Cleaning waste separator