Temperature correctionThe assayed values are based on an 60 minutes ESR, with dilution and temperature correction. Therefore, the measured ESR value should be compared with the expected value using temperature correction. The calculation of a 30 minute measurement to a 60 minute ESR result with temperature correction influences the QC result due to rounding. See chapter QC Results for more information.
|
||||||