Maintenance

The Interrliner is an analyzer that operates with considerable amounts of whole blood virtually undiluted, and stores it in a pipette for one hour. For this reason instrument maintenance is of the utmost importance.

To maintain the maximum reliability of the instrument, the maintenance procedures must be strictly followed. All procedures are based on a number of samples.

Maintenance levels

Work instruction

Daily

Daily

Weekly

Weekly maintenance

Level 4

Level 4 maintenance Every 7500 samples

Level 3

Level 3 maintenance Every 23000 samples

Level 2

Level 2 maintenance Every 46000 samples

Level 1

Level 1 maintenance Every 91000 samples

Two yearly

Every two years

Note: Numbers are based on 5 days week with 350 samples per day.

bio-hazard

WARNING!!!
Always be aware of the danger of infection, especially during maintenance. Take appropriate precautions. There is blood involved and therefore a BIO HAZARD

In this chapter

Daily

Weekly

Level 4 maintenance

Level 3 maintenance

Check or replace sample probe or outer needle

Reagents installation

Cleaning Touch screen display