Explain QC Settings assay range

Use default assay range

ON (=Default setting): The default assay range which is predetermined for the current batch of Starrsed Control is used. These ranges cannot be changed in this option.

OFF: Use custom assay range.

Use custom assay range

ON: The lab can establish their own acceptable ranges, both ranges can be set from a minimum of +/- 2 mm to a maximum of +/- 15 mm.

OFF: Use default assay range.

Note: It is advised to use the default assay ranges. Use caution when setting the custom ranges. A too narrow range may cause unjustified rejection of QC sample results, subsequent rejection of patient results and undue burden on maintenance.

Return to QC Settings (QC Settings, QC Settings) creen.