Rinse head replacement
Switch Starrsed OFF.
Rinse head disassembly:
- Loosen the plastic grub screw.
- Disconnect the two tubes from the rinse head.
- Hold the rinse head solenoid core tight and turn the rinse head anti clockwise.
- Remove the rinse head from the instrument.
- Remove the black protection bellow.
- Remove the piercing needle, use a 1.5mm hex driver.
![rinse_nozzle_wrong_old](rinse_nozzle_wrong_old.jpg)
|
![rinse_nozzzle_good_old](rinse_nozzzle_good_old.jpg)
|
Wrong position. Piercing pin to high
|
Correct position
|
Rinse head assembly:
- Re-install the rinse head, hold the solenoid core while turning the rinse head clockwise.
- Push the rinse head core upwards, the rinse head should touch the V shaped rubber ring. Make sure that the core reaches the end of his stroke!
- Tighten the plastic grub screw.
- Winding the rinse head clockwise will cause a bigger gap between the V-ring and top of the rinse head.
|