Measure sensor
- Check Measure sensor (In Service mode): When the sensor is not engaged with the pipette, the Value must be within the following limits: MS .50 ±10.
Press the button . The pipette currently at the measure position will be measured. The results are displayed in graphical form. Raw data is also stored on the D: drive (D:\MeasureTest.txt).
![take out pipette bza](measure_start_ok.jpg)
|
Measure head start position correct
|
![take out pipette bza](measure_start_wrong.jpg)
|
Measure head start position wrong
|
NOTE: Clean sensors first before executing this function. When a test pipette is installed at the measuring position the result of the test pipette is displayed in the field "ESR (mm)".
|